Turbimax CUS50D
Technical data
Hauser
43
13
Technical data
13.1 Input
13.1.1 Measured variables
• Turbidity
• Absorption
• Solids content
• Product loss
• Temperature
13.1.2 Measuring range
Application
Specified operating range
Maximum operating range
Absorption factory calibration
0.000 to 5.000 AU or
0.000 to 10.000 OD
Factory calibration for
formazine
40 to 4,000 FAU
10000 FAU
Application: Kaolin
0 to 60 g/l
500 g/l
Application: Sludge
0 to 25 g/l
500 g/l
Application: Auto sludge
0 to 25 g/l
500 g/l
Product loss
0 to 100 %
1000%
Measuring range with solids content:
For solids, the achievable ranges depend very much on the media that are actually
present and may differ from the recommended operating ranges. Extremely
inhomogeneous media may cause fluctuations in measured values, thus narrowing the
measuring range.
13.2 Performance characteristics
13.2.1 Reference operating conditions
20 °C (68 °F), 1013 hPa (15 psi)
13.2.2 Maximum measured error
Absorption
0.5 % of the upper range value (corresponds to ± 50 mOD)
Formazine
10 % of the measured value or 10 FAU (the greater value applies in each case)
Kaolin
5 % of the upper range value; applies to sensors that are calibrated for the observed measuring
range