Installation
Turbimax CUS50D
20
Hauser
Installing with Flexdip CYA112 immersion assembly and Flexdip CYH112 holder
A0036839
15
Installing with immersion assembly
The installation angle is 0°.
The arrow indicates the flow direction; it
runs from the 10 mm (0.39 in) path to the
5 mm (0.2 in) path.
If the sensor is used in open basins, install
the sensor in such a way that air bubbles
cannot accumulate on it.
5.3
Mounting the compressed air cleaning unit
‣
Fit the compressed air cleaning unit onto the sensor head to the end stop. The nozzle of the
compressed air cleaning unit must be located on the side of the wider 10 mm (0.4 in)
measurement gap.
5.4
Post-installation check
Put the sensor into operation only if the following questions can be answered with "yes":
• Are the sensor and cable undamaged?
• Is the orientation correct?
• Has the sensor been installed in the process connection, and does not suspend freely from
the cable?