TopClean S CPC 30
6 Commissioning
Hauser
89
Calibration
redox relative
"Calibration relative"
For calibration, two tanks are filled with a sample of the medium. The contents of the first
tank are detoxified and called calibration solution 1.
The contents of the second tank are left unchanged and are called calibration
solution 2.
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
CC
Calibration with
abs. calibration
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
–
CCD1 / CCD4
Immerse:
Electrode in buffer 1
Handling instructions
Immerse the electrode in buffer 1 / 2 (detoxified
sample, see above) and then press "E".
–
CCD2 / CCD5
20 %
(0 ... 30 %)
Enter buffer
Enter the relative redox value of buffer 1 / 2 (detoxified
sample) in percent.
E2
CCD3 / CCD6
Time: 10 s
mV 1:
225
Checks the stability of the calibration
Wait until the measurement is stable:
Time does not count down,
mV value no longer flashes,
Display "Measured value stable"
Then press "E".
–
CCD7
Invalid calibration value
Note in display:
Error message if the offset is too large.
–
CCD8
Accept
Reject
Recalibrate
End of calibration
Accept: Pressing "E" accepts the new calibration data.
Reject: The data is not accepted, a recalibration won’t
be started.
Recalibrate: The data is rejected and a new calibration
will be started.
E1
CCD9
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that
measurement can take place?
–