8 Trouble-shooting
TopClean S CPC 30
106
EHauser
Behaviour with NAMUR setting
8.2.3
Response of contacts to power failure
In the "Set up 1" menu
➞
"Contacts", you can define the contacts as NC contacts or NO
contacts (s. page 38). In the case of a power failure, the contacts will act according to
the setting you make.
8.2.4
Assembly behaviour
Instrument status
Alarm relay
Maintenance
relay
Function
check
Limit value /
Controller
Normal operation
Picked-up
(fail-safe
behaviour)
Appropriate
configuration and
operating status
Failure
Dropped out
Maintenance required
Picked up
Function check
Picked up
Voltage-free
Dropped out
Dropped out
Problem
CPM 153 behaviour
CPG 30 behaviour
Assembly behaviour
CPM 153 voltage
failure during
measurement
No function
CPM 153 voltage
failure during
maintenance
No function
CPG 30 voltage
failure during
measurement
Error message E012
All valves closed
Can be pushed out of
the process.
CPG 30 voltage
failure during
maintenance
Error message E012
All valves closed
Stays in
service position
Air failure during
measurement
Error message
Can be pushed out of
the process.
Air failure during
maintenance
Error message E027
Stays in
service position
Air failure during
autoclean
Error message E027
Stays in
service position
Water failure during
measurement
Error message E028
Water failure during
maintenance
Error message E028