Commissioning
120
Relative data input (only for redox relative)
Entry of two % calibration points to each of which one mV value is assigned.
Press
Ï
to start calibration.
Relative calibration (only redox relative)
For calibration, two tanks are filled with a sample of the medium. The contents of the first tank are
detoxicated and called calibration solution 1.
The contents of the second tank are left unchanged and are called calibration solution 2.
Press
Ï
to start calibration.
Function
Options
Info
Note
Calibration with enter data rel.
The type of local calibration selected in the
calibration settings is displayed.
Calibration points
Activation and subsequent entry
• 0 ... 30%
• Voltage
-1500 ... +1500 mV
• 70 ... 100%
• Voltage
-1500 ... +1500 mV
Factory setting
20 %
Voltage: 0600 mV
80 %
Voltage: -600 mV
In this field, create two measured value pairs
(pair 1 and pair 2).
Measured value pair 1 in the range 0...30 %:
Assign, for example, the voltage 0600 mV to
the percentage value 20 %.
Measured value pair 2 in the range 70...100 %:
Assign, for example, the voltage -0600 mV to
the percentage value 80 %.
The settings made are effective immediately
after confirming with
F
.
Note
Offset too high
Error message if the entered offset leaves the
maximum range.
Calibration
Options
• Accept
• Cancel
• Repeat calibration
Ending calibration
Accept:
When you confirm with
F
, the new
calibration data are accepted.
Cancel:
The data are not accepted. Calibration
is not repeated.
Repeat calibration:
The data are rejected and
calibration is repeated.
Sensor communication
Waiting for sensor response ...
(only available for digital sensors with
Memosens technology.)
The transmitter transmits calibration data to the
sensor.
Note
• Data saved
• Data NOT saved
(only available for digital sensors with
Memosens technology.)
Indicates whether the calibration data could be
saved in the sensor. Calibrate the sensor again if
the process of saving the data failed.
Note
Electrode in medium?
Ensure that the electrode is in the medium
again so measuring can take place.
Function
Options
Info
Note
Calibration with calibration rel.
The type of local calibration selected in the
calibration settings is displayed.
Handling instructions
Immerse:
Electrode in buffer
Immerse the electrode in the detoxicated
sample.
Confirm with
F
.
% value buffer
0 ... 30%
Factory setting
20 %
Enter the relative redox value of the
detoxicated sample.
Confirm with
F
.