Operation
TOCII CA72TOC
30
Hauser
5.3
Operation during measuring mode
The analyzer has three operating modes:
• Measuring mode
• Service mode
• Programming mode
The measuring mode is fully automated. Manual intervention on the part of the user is not possible.
a0012443
Fig. 11:
Display during measuring operation
1
Time
2
Load curve of the past 6 hours
3
Timeframe
4
Measured value
5
Measured value of the IR detector
5.3.1
Recording mode
Pressing the
U
key during measuring operation takes you to the recording mode. Using the arrow
keys, scroll between the measured values recorded:
•
V
1 day earlier
•
W
1 day later
•
U
2 hours earlier
•
T
2 hours later
Once you have selected the required period, press the Enter key.
Recording period
• 14 days for one-channel operation
• 7 days for two-channel operation
5.3.2
Spot view
The spot view is activated in the recording mode by pressing the Enter key.
The following information is displayed:
• Load curve
• Measured value
• Date
• Time
A small cursor on the display indicates where you are currently in the load curve.
TOC
mg/l
IR = 334 ppm
234
5
00
6
00
7
00
8
00
9
00
10
00
11
00
1000
500
11:39:04
1
2
3
4