Commissioning
TOC Analyzer CA79
32
Hauser
4. Press OK in the Login window.
Pressing the Enter character on a connected keyboard or moving the mouse to the
next line will produce a login error.
A pop-up window appears once the various pages are closed. Wait for approx. 30 seconds
to ensure that all the data are saved.
Once you close the software program and shut down Windows, you can switch off the
device with the mains switch.
8.3.8
Saving measured data
The measured data should be backed up at regular intervals. A USB hub with at least 4
ports, a mouse, a keyboard and a USB stick with at least 8 GB of space are needed for the
backup.
1. Shut down the system entirely .
The main menu is displayed.
2. Open the
Viewer
menu.
3. Press the
Online
key to select the online data.
4. Select the
Table
tab.
5. Press the
Export csv
key.
The file manager opens.
6. Copy the folder to be saved and save it to the connected USB stick.
8.3.9
Available options
Interval mode editor (order option)
A sequence of up to 8 combinations of port inputs between the sample, input 2 and input
3 can be created in this editor.
The interval editor can be opened via the
Interval list
key (
Settings
menu ->
Options 1
tab).
A0046973
23
Settings menu, Options 1 tab
If the
Calibr. single port
option is selected, input 1 can be used for the calibration or
SST without having to disconnect the cables.