Commissioning
Teqwave F/I Modbus TCP
44
Hauser
2. Settings menu → "Application parameters" → "Diagnosis" → "Change in → Select
measured variable
3. Settings menu → "Application parameters" → "Diagnosis" → "Process disturbance" →
"Switch point"
Navigation using the Viewer
1. Menu "Teqwave Transmitter" → "View filter" → "Filter options" and "Filter actions"
2. Menu "Teqwave Transmitter" → "View filter" → "Change in" → Select measured variable
3. Menu "Teqwave Transmitter" → "View filter" → "Process disturbance" → "Switch point"
Parameter overview with brief description
Parameter
Procedure
Selection/input
Factory setting
View options
(transmitter)
Filter options/filter
actions
(Viewer)
Select filter options and arrangements to enable or disable them.
Filter options:
• Enable measuring
• Enable calibration
• Stationarity
Filter actions:
• Hide measured value
• Background coloring
• Hold measured value
The "Enable
measuring" and
"Enable calibration"
options are enabled.
The "Stationarity"
option is disabled.
Change in
concentration 1 to 2
[unit]
Specify the limit value for the maximum permissible speed at which
concentration can change. Enable or disable the function using the
button.
If the measuring unit is changed, the unit is changed according to
the "change in" setting, but the value entered is not converted.
0.01 to 100[unit]
100 [unit]
The function is
disabled.
Change in
temperature °C/min
(°F/min; °K/min)
Specify the limit value for the maximum permissible speed at which the
temperature can change. Enable or disable the function using the button.
If the measuring unit is changed, the unit is changed according to
the "change in" setting, and the value entered is converted.
0.01 to 100 °C/min
(K/min) or 0.018 to
180 °F/min
1.5 °C/min
The function is
enabled.
Change in speed of
sound (m/s)/s
Enter the limit value for the maximum permissible speed at which the
speed of sound can change. Enable or disable the function using the
button.
0.01 to 100 (m/s)/s
100 (m/s)/s
The function is
disabled.
Change in density
[unit]
Enter the limit value for the maximum permissible speed at which the
density can change. Enable or disable the function using the button.
If the measuring unit is changed, the unit is changed according to
the "change in" setting, and the value entered is converted.
0.01 to 100 (kg/m3)/s or
(g/l)/s 0.00001 to 0.10
(g/cm3)/s
100 (kg/m3)/s
The function is
disabled.
Change in dispersion
1/s
Enter the limit value for the maximum permissible speed at which the
dispersion can change. Enable or disable the function using the button.
0.01 to 100 1/s
100 1/s The function
is disabled.
Switch point
Set the trigger value for process disturbances (dispersion).
Transmitter
Enable or disable the function using the Enable / Disable button.
Viewer
Enable or disable the function using the
Process disturbance
button.
• If the speed of sound and temperature are used for concentration
measurement, do not exceed the value of 1. If the density is used,
apply 0.25 as the switch point.
• If the measured dispersion is greater than the configured switch
point, the measuring device does not display the concentration
any more.
0.01 to 1
0.3
The function is
disabled.
10.5 Advanced settings
The measuring device supports the performance of a field calibration.