Technical data
42
10.4
Performance characteristics
10.5
Environment
10.6
Process
10.7
Mechanical construction
Time between two measurements
t
mes
= reaction time + rinse time + waiting time + rinse again time + filling
time + sampling time + reagent refusal time (min. waiting time = 0 min)
Maximum measured error
±2 % of measuring range end
Measuring interval
t
mes
to 120 min
Reaction time
2 minutes
Sample requirement
15 ml (0.004 US.gal.) per measurement
Reagent requirement
1 x 0.21 ml (0.000055 US.gal)
0.91 l (0.24 US.gal) per reagent per month with 10 minute measuring
interval
Calibration interval
0 to 720 h
Rinse interval
0 to 720 h
Rinse time
selectable from 20 to 300 s (standard = 60 s)
Rinse again time
30 s
Filling time
22 s
Maintenance interval
6 months (typical)
Servicing requirement
15 minutes per week (typical)
Ambient temperature
5 ... 40 °C (41... 104 °F), avoid strong fluctuations
Humidity
below the condensation limit, installation in usual, clean rooms
outdoor installation only possible with protective devices (customer
supplied)
Ingress protection
IP 43
Sample flow rate
min. 5 ml (0.0013 US.gal.) per min
Consistance of the sample
low solid content (< 50 ppm)
Sample inlet
pressureless
Design, dimensions
see chapter "Installation"
Weight
GFK housing
approx. 28 kg (61.7 lb)
Stainless steel housing
approx. 33 kg (72.8 lb)
Materials
Housing
Stainless steel 1.4301 (AISI 304) or GFK
Front windows
Plexiglass
®
Endless hose
C-Flex
®
, Norprene
®
Pump hose
Tygon
®
, Viton
®
Valves
Tygon
®