Trouble-shooting
Hauser
45
9.3
Process errors without messages
Error
Possible cause
Tests and / or corrective measures
Measured values always
the same
Reagents contaminated
or aged
The dichromate solution R2 must be orange to brown in color. It should not have a green hue.
The sulfuric acid R1 must be clear and colorless.
No sample, no reagents
Make sure sample and reagents are supplied, check level monitor and clean if necessary
System blocked
Flush with 12.5% bleaching lye.
Measured values
inaccurate
Incorrect concentration
of standard
Check the concentration in the laboratory. Adjust the standard accordingly ("PARAMETER ENTRY",
"Calibration solution") or replace the standard.
Reagents contaminated
or aged
The dichromate solution must be orange to brown in color. It should not have a green hue.
Reagent blank value too
high
After replacing the reagents, carry out an offset calibration and then a calibration (CONFIGURATION,
"Offset calibration")
Wrong cell
Check the setting in the CONFIGURATION menu, "Photometer"
Sample suction time too
short
Increase the number of rinse cycles (CONFIGURATION, "Rinse cycles").
Matrix effects
(substances that interfere
with the photometric
method)
Detect interfering substances (see Technical Information, "Measuring principle"),
possibly use sample conditioning
Filter life too long
Take a counter sample at the analyzer inlet and check the concentration in the laboratory. If there is no
deviation to the analyzer measured values, replace the ultrafiltration modules or reversible flow filters more
frequently.
System blocked or
contaminated
Flush with 12.5% bleaching lye (monthly maintenance)
Dosing
Replace pump hoses.
Cell dirty
First clean with 2% ammonium hydroxide solution and then rinse with water.
Counter sample in the
laboratory delivers
deviating measured
values
Sample aging
Shorten time between sampling and analysis.
Measured value transfer
defective
Wrong size for analog
output
Check the setting (CONFIGURATION, "Analog output 1" or " 2").
Incorrect measuring
range
Adjust measuring range (PARAMETER ENTRY, "Measuring range")
Background noise
Check electrical cables for interference from strong sources of induction.
Analyzer will not switch
on
No power
Check the electrical connection and ensure power supply.
Fuse
Replace fuse F1 (slow-blow 0.5 A)
Analyzer is running but
display is incomplete or
off
Initialization failed
Switch the analyzer off and after approx. 30 seconds back on.
Pumps fail to run
Leak
See "Spillage error" error message
Leak sensor bypassed
Interrupt the contact between the two leak sensors (terminals 67-66)
Fuse
Check all fuses and replace if necessary.
Pump defective
Service
Measurement does not
start
Leak at photometer
Service
"Measurement" display
flashes
Time of 1st
measurement not
reached
The date must be between 01.01.1996 and the current date.
Interval not expired
Change parameter setting.