7.7
Placing out of service
You must place the analyzer out of service before shipping or before longer operation breaks (more
than 2 days).
CAUTION
!
Residues of chemicals can cause injuries
►
Thoroughly rinse all of the lines of the measuring system with clean water.
To place the analyzer out of service, proceed as follows:
1.
Remove the reagent and standard hoses from the canisters and immerse them in a tank
containing clean water.
2.
In case of using a continous sample supply: disconnect the sample feed and immerse the hose
in the tank too.
3.
Switch pump P2 on for 3 min (see "Service" section in "Operation" chapter).
4.
Empty the reactor by switching the valves to passage for chromate containing waste (V1: S,
V4: a). Close aagin the valve (V1: P, V4: z).
5.
Switch successively:
a. V1: P, V2: S, V3: H and pump P1 on for 1 min.
b. V1: S, V2: S, V3: P and pump P1 on for 1 min.
c. V1: S, V2: O, V3: P and pump P1 on for 1 min.
d. V1: P, V2: S, V3: P and pump P1 on for 1 min.
6.
Remove the hoses from the water and repeat steps 3 and 5. Allow the pumps to run until the
hoses are completely dry.
7.
Flush sampling hoses with clean water and then with compressed air, in order to completely
empty the hoses.
8.
Remove the valve hoses from the valves.
9.
Remove the load from the pump hoses by removing the hose cassette from the bracket below.
Keep opened reagents and standards in a refrigerator. Observe the shelf-life.