Maintenance
Stamolys CA71CODcr
38
Hauser
7.3.2
Back-flush pump
Removing the old hose
1.
Flush the old hose with water first and then with air to empty it (see SERVICE).
2.
De-energize the back-flush pump.
3.
Loosen the screws 1 to 3 of the safety cover and remove the cover.
4.
Remove the old hose.
Installing the new hose
1.
Install the new hose. Fix the white plastic clips in the guidances.
2.
Grease the new pump hose with silicone grease.
3.
Attach the safety cover again and fix it by the screws 1 to 3.
4.
Energize the back-flush pump.
7.4
Setting the hose cases
The analyzer is supplied with preset hose cases.
You must only reset these cases if you discover that little or no medium is pumped after a hose has
been replaced.
1.
Place a beaker with distilled into the device and place the hoses for sample and reagent R2 into
the beaker.
2.
Flush the hoses (sample and reagent R2) with water first and then with air to empty them (see
SERVICE).
3.
Disconnect the pump hoses (item 2) coming from pumps P1 and P2 from the valve manifold:
a0017281
Fig. 22: Back-flush
pump