Maintenance
Stamolys CA71CODcr
34
Hauser
7
Maintenance
CAUTION
!
You must not carry out any procedures not listed in the following chapters, yourself.
►
Not listed works must only be carried out by the service.
►
Wear your personal protective clothing when carrying out any maintenance work.
If you use components for maintenance or modification, which are not approved by
Hauser, the warranty and the certified conformity of the device will become void.
7.1
Maintenance schedule
All maintenance duties that have to be carried out during normal operation of the analyzer are
explained below.
If you are using a sample conditioning unit, e.g. CAT430, coordinate the maintenance work
required for it with that of the analyzer. For this, read the maintenance chapter in the respective
operating instructions.
Whenever working on the reagent hoses, the hoses must be disconnected from the canisters,
in order to prevent contamination of the reagents. Empty the reagent hoses via the service
menu.
7.2
Replacing reagents
CAUTION
!
Corrosive chemicals and other risks
Eye or skin injuries or crushes possible
►
There is a danger of crushing limbs at doors, inserts and pump heads.
►
Refer to the warning instructions in the safety data sheets when handling reagents or cleaning
solutions. Wear protective clothing, gloves and goggles.
►
Wear protective clothing, gloves and goggles.
►
Never add water to reagents. Reagents containing acids may splash and heat may build up.
If you store the reagents correctly (in the dark, not over 20
˚
C) they will be stable for minimum
12 weeks from the date of manufacture (batch number). When this period of time has expired, the
Period of time
Duty
Note
Weekly
– Check and note calibration factor (for service
purposes)
– Move valve hoses into their position and
spray with silicone (extends the service life).
CONFIGURATION
Every 2 weeks
– Replace reagents
– see chapter "Replacing reagents"
Monthly
– Flush sample line hose system with
pressurized water (disposable syringe), check
and replace reagents if necessary
– Grease pump hoses with silicone grease
– Check sample collector for fouling and clean
it if necessary
– Remove the cassette of the
sample pump.
– Connect the one-way syringe instead of the
sample inlet.
Every 6 weeks
– Rotate pump hoses
Every 3 months
– Cleaning the drain lines:
Flush all hoseswith sample for at least
30 minutes
– Replace pump hoses
– Cleaning of the filter mat of the housing fan
– SERVICE:
V1: P, P1: g, P2: s, V2: S, V3: P
Add solution to sampling connection
– s. chapter "Replacing the pump hoses"
– s. chapter "Cleaning the filter mat of the
housing fan"
Every 6 months
– Replace valve hoses
– see chapter "Replacing valve hoses"