Operation
Stamolys CA71CODcr
28
Hauser
5.2.9
DATA STORAGE
5.3
Timers (version with back-flush pump only)
When the pump is switched on the T1 fields are lit or flash red and yellow on the display, as does
the OP message on the bottom left of the screen.
The red number at the top displays the switch-on time that has already elapsed while the yellow
value at the bottom displays the target value for the switch-on time. All values are full seconds.
When the pump is switched off the T2 fields are lit or flash, as do the actual and target values for
the switch-off time. The OP message is not displayed.
The optimum switch cycle and the time values are normally already preset when the device is
delivered from the factory. However, you might need to alter the settings if the process conditions
change or a timer has to be replaced.
5.3.1
Time range
The correct time range for the switch-on and switch-off time is between 1 and 9999 seconds. It is
normally set at the factory and should not be changed. To change the time ranges, you first have to
remove the timer.
Selection
Display
Info
Measured values
The data memory contains the last 1024 concentration
measured values with date and time. If there are no
values available, "Empty set" appears.
Browse through the data sets by pressing the
V
and
W
keys.
Serial output
You can output
all
data sets (in ASCII-format) via the
serial interface. For this, the receiving end (PC) must be
configured like this: 9600, N, 8, 1.
To send data, the receiving end (PC) must send the ASCII
character 81 ("Shift", "D").
Clear data
This deletes
all
data sets.
a0015645
Fig. 17: Timer