Stamolys CA71CODcr
Installation
Hauser
9
3.2.2
Connecting the sample line
3.3
Installation instructions
3.3.1
Analyzer
To install the analyzer at the intended location, proceed as follows:
1.
Mount the analyzer and secure it to a vibration-free wall using screws (Ø6 mm / 0.24"). The
fastening clips on the housing make sure that the necessary distance to the wall is observed for
ventilation purposes.
For the installation dimensions, please see the previous chapter.
2.
Using a spirit level, check that the cabinet is hanging straight. Only in this way can any air
bubbles present escape from the cell.
3.
Fit the edge covers (with GFR housing only).
4.
Lay the drain pipe for the reaction products. Where possible, use solid pipes (PE, internal
diameter 1" with 3% incline).
5.
Route the drain hoses through the device floor to the outside.
a0016672
Fig. 3: Collecting
vessel
1
Connection for level measurement
2
Ventilation
3
Sample supply to analyzer, connection
for hose with ID3.2mm(0.13")
4
Connecting cable of electric agitator
5
Sample supply, connection for hose with
ID3.2mm (0.13")
6
Sample overflow, connection for hose
with ID10mm (0.39")
a0016671
Fig. 4: Dimensions:
collecting vessel and base plate
Collecting vessel
(at analyzer, with or without level measurement)
Connection
hose ID 3.2 mm (1/8 inch)
Customer collecting vessel
Connection
hose ID 1.6 mm (1/16 inch)
Max. distance from collecting vessel to analyzer
1 m (3.3 ft)
Max. height difference from collecting vessel to analyzer
0.5 m (1.6 ft)