Stamolys CA71HY
Operation
Hauser
27
!
Note!
• Always coordinate calibration and flushing.
• Carry out flushing with standard cleaners approx. 3-4 hours
before
the next calibration.
• Flushing with special cleaners (e.g. hydrochloric acid or ammonia) has a lasting effect on
calibration. Therefore, carry out this cleaning
3-4 hours before
calibration. Then rinse with
deionized water afterwards.
5.2.5
LANGUAGE
The following languages are available:
• Deutsch
• English
• Français
• Espanol
• Suomi
• Polski
• Italiano
5.2.6
ERROR DISPLAY
!
Note!
• This menu is a "Read-Only-Menu".
• You can find the individual error messages, their meaning and solutions to problems in the
"Troubleshooting instructions" chapter.
• If at least one error message is present, the signal output is set to "fault".
• Causes of faults are queried with every measurement. If an error which occurred previously no
longer exists, the message is automatically deleted. If this does not happen automatically, error
messages can be deleted by quickly switching the analyzer off and back on again.
1. flushing
01.01.96 00:00 to 31.12.95 23:59
Time of 1st flushing (DD.MM.YY, time hh.mm). After
each change the instrument does not wait for the flushing
interval. If the flushing is to start immediately, set the
time in the past.
– Set the time to 4:00 to track the effects of flushing in
the curve.
– If you have started flushing manually, you should
redefine the time of the 1st flushing because the
interval is calculated starting from the last flushing.
Flushing interval
0 to 720 h
48 h
Time between two flushings.
The "0 h" setting means that cleaning no longer takes
place.
Flushing hold on
0 to 60 s
60 s
Dwell time of flushing solution in the
pump-mixer-photometer line.
Recommended: 30 to 60 s.
1)
Option 0 or 4 mA: see "CONFIGURATION".
2)
All channels set to 0 means that the channel is selected externally. All channels set to 1 means alternating, beginning with channel 1.
Menu item
Range of adjustment
(factory settings in bold)
Display
Info
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...