Wiring
Stamolys CA71HY
20
Hauser
!
Note!
Condition met means:
• AV 1: concentration > alarm value 1
• AV 2: concentration > alarm value 2
• Fault: error occurred
Contacts AV 1, AV 2 and fault are only affected during automatic operation.
4.4
Serial interface
The results (measured value+engineering unit) are output in the "Data memory-Measured values"
menu.
The calibration results (measured value+engineering unit) are output in the "Data
memory-Calibration factors" menu.
!
Note!
• A null modem cable is required (not a crossed one).
• The interface does not have to be configured at the analyzer.
The following commands can be sent from the PC in order to read out data:
• "D" = Data memory-Measured values
• "C" = Data memory-Calibration factors
• "S" = Setup (parameter entry, configuration...)
• "F" = Frequency (current)
Channel 1/2
measurement end
A:
R:
22
22
-
-
20
21
A:
R:
22
22
-
-
21
20
22
-
21
A = NO configured
R = NC configured
Terminal connection for
condition fulfilled
Terminal connection for
condition not fulfilled
Terminal connection for
power off
RS 232 of CA 71
COM 1/2 at PC
SUB-D, nine-pin
Function
Function
SUB-D, nine-pin
3
TxD
RxD
2
2
RxD
TxD
3
8
CTS
RTS
7
CTS
8
5
GND
GND
5
Software protocol:
Output format:
9600, N, 8, 1
ASCII
Содержание StamoLys CA 71 HY
Страница 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Страница 48: ...Appendix Stamolys CA71HY 48 Endress Hauser 11 Appendix 11 1 Operating matrix One channel version a0001907 en...
Страница 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Страница 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Страница 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Страница 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Страница 59: ......
Страница 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...