3
–
3
Lifting/carrying the analyzer
Due to the analyzer’s size and weight (48.6 Kg., or 103 lbs.), Hauser
recommends the use of a forklift, pallet jack, etc. to lift and/or move the analyzer.
Before removing the analyzer from the crate, move the crate as close as possible
to the final location.
If the analyzer is to be lifted by hand, designate multiple individuals to lift by the
mounting brackets, and distribute the weight among personnel to avoid injury.
Refer to Appendix A for drawings and specifications of the analyzer.
Mounting the analyzer
The SS500XP/SS2000XP analyzer is manufactured for wall or Unistrut
(or
equivalent) metal framing installations. Refer to the layout diagrams in Appendix A
for specifications.
To mount the analyzer
1.
Select a suitable location to mount the analyzer. Choose a shaded area
or use an optional analyzer hood (or equivalent) to minimize sun
exposure.
2.
Locate the mounting holes on your unit. Refer to Figure A–1 on
3.
For wall installations, mark the centers of the holes in the mounting
brackets to the mounting surface.
4.
Drill the appropriate size holes for the bolts you are using
(Hauser recommends 2 in. bolts at a minimum).
5.
Hold the analyzer in place and fasten with the top screws.
6.
Repeat for the bottom mounting holes.
When mounting the analyzer, be sure not to position the
instrument so that it is difficult to operate adjacent devices. Allow
1 meter (3 feet) of room in front of the analyzer and any switches.
It is critical to mount the analyzer so that the inlet and outlet lines
reach the inlet and outlet connections on the chassis while still
maintaining flexibility so that the sample lines are not under
excessive stress.
Hauser analyzers are designed for operation within the
specified ambient temperature range of -20 °C to 50 °C (
-
4 °F to
122 °F). Intense sun exposure in some areas may cause the
analyzer temperature to exceed the maximum.