TDLAS Gas Analyzer
XA02744C
Hauser
47
If the membrane filter is wet (liquid or contaminants are detected on the fil
-
ter):
1.
Drain any liquids and clean with isopropyl alcohol.
2.
Clean any liquids or contaminants from the base of the membrane sep
-
arator.
3.
Replace the filter and the O-Ring.
4.
Place the cap onto the membrane separator and tighten with a channel
lock wrench.
5.
Check upstream of the membrane for liquid contamination and clean
and dry out before opening the shut-off valve at the sample tap.
Service
For Service in your location, please refer to our website (https://
www.endress.com/contact) for the list of local sales channels.
For the SS500e, SS2000e, and SS3000e analyzers, replace the
four screws and tighten the base.