TDLAS Gas Analyzer
XA02744C
Hauser
25
Equipment
Installation
The information in this section is related to safety considerations for equip
-
ment installation.
Lifting/carrying the analyzer
Due to the analyzer’s size and weight (48.6 Kg., or 103 lbs.), Hauser
recommends the use of a forklift, pallet jack, etc. to lift and/or move the ana
-
lyzer. Before removing the analyzer from the crate, move the crate as close
as possible to the final location.
Electronics with
sample cell
dimensions
565 mm H x 413 mm W x 222 mm D (22.25 x
16.25 x 8.75 inches)
Weight approximate
46.8 kg (103 lbs)
Sample cell
dimensions
438 mm H x 108 mm W (17.3 x 4.3 inches)
Number of Sample
Cells
1
Area classification
Certification
cCSAus Class I, Div 1, Groups B, C, D, T4
Hauser Class I Division 2 analyzers use a non-
incendive protection method, and as such all portions of the
local installation codes apply. The maximum allowed
inductance to resistance ratio (L/R ratio) for the field wiring
interface must be less than 25 µH/Ω. The maximum total
loop capacitance shall be 0.27 microfarads.
The safety of the analyzer is the responsibility of the installer
and the organization he/she represents.
Configurations requiring optional accessories, e.g., probe
assemblies, with specific characteristics must meet
manufacturer specifications.
Hauser analyzers are designed for operation within
the specified ambient temperature range of -20 °C to 50 °C (-
4 °F to 122 °F). Intense sun exposure in some areas may
cause the analyzer temperature to exceed the maximum.