Specifications
Hardware Installation and Maintenance Manual
A
–
3
Table A–2
2-Pack: Analyzer A (H
2
S) / Analyzer B (H
2
O) specifications
Performance
Concentration
1
1. Consult factory for alternative ranges.
Refer to calibration report
Repeatability
Refer to calibration report
Measurement Update Time
2
2.
Total system response is dependent on flow rate and sample system volume
.
H
2
S
: <5 seconds
H
2
O
: 1 second
Periodic Scrubber Cycle Duration
90 seconds
Application Data
Environmental Temperature
Range/ Sample Cell Temperature
Range
–20 °C to 50 °C (–4 °F to 122 °F) -
Standard
–10 °C to 60 °C (14 °F to 140 °F) -
Optional
Heated SCS Enclosure
Temperature
50 ± 1 °C
Environmental Relative Humidity
5% to 95%, Non-condensing
Altitude
Up to 2000 m
Sample Inlet Pressure
1.5 to 4 barg (20 to 50 PSIG) to enclosure inlet
Cell Pressure Range
800 to 1200 mbar -
Standard
950 to 1700 mbar -
Optional
Sample Flow Rate
H
2
S
: 8 SCFH (4 LPM)
H
2
O
: 2 SCFH (1 LPM)
Recommended Validation
H
2
S
: Binary cal gas bottle
H
2
O
: Bureau of Mines Chilled Mirror, portable
TDL or binary cal gas
Electrical & Communications
Input Voltage
3
3. Supply voltage not to exceed
10% of nominal. Transient over-voltages according to
Over Voltage Category II.
Analyzer
:
120 VAC or 240 VAC
±
10%, 50/60 Hz, 60 W
maximum
18 VDC to 24 VDC, 1.6 A maximum
SCS
: 120 VAC or 240 VAC
±
10%, 50/60 Hz, 200
W maximum
Current (Analyzer)
3 A maximum at 120 VAC, 1.5 A maximum at
240 VAC
Current (SCS enclosure heater)
2 A maximum at 120 VAC, 1 A maximum at
240 VAC
Contact Rating (Inductive Load)
250 V, 3 A N.O. contact, 1.5 A N.C. contact
24 V, 1 A N.O. and N.C. contact
Содержание SS1000
Страница 2: ......
Страница 4: ......
Страница 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Страница 121: ...www spectrasensors com contact 4900002256 Rev B ...