Troubleshooting & Maintenance
Hardware Manual
B
–
5
Tools and Supplies:
•
Lens cleaning cloth (Cole Parmer
®
EW-33677-00 TEXWIPE
®
Alphawipe
®
Low-Particulate Clean Room Wipes or equivalent)
•
Reagent-grade isopropyl alcohol (ColeParmer
®
EW-88361-80 or
equivalent)
•
Small drop dispenser bottle (Nalgene
®
2414 FEP Drop Dispenser
Bottle or equivalent)
•
Acetone-impenetrable gloves (North NOR CE412W Nitrile Chemsoft™
CE Cleanroom Gloves or equivalent)
•
Hemostat (Fisherbrand™ 13-812-24 Rochester-Pean Serrated
Forceps)
•
Bulb blower or dry compressed air/nitrogen
•
Torque wrench
•
Permanent ink marker
•
Non-outgassing grease
•
Flashlight
To clean the mirrors:
1.
Power down the analyzer following the procedure outlined in
“To
power down the analyzer”
in the Firmware Manual.
2.
Isolate the sample conditioning system
(
SCS) from the process
sample tap.
The sample cell assembly contains a low-power, 20 mW MAX, CW
Class 3b invisible laser with a wavelength between 800-3000 nm.
Never open the sample cell flanges or the optical assembly unless
the power is turned off.
The sample cell assembly contains a low-power, 20 mW MAX, CW
Class 3b invisible laser with a wavelength between 800-3000 nm.
Never open the sample cell flanges or the optical assembly unless
the power is turned off.
All valves, regulators, switches, etc. should be operated in
accordance with site lock-out/tag-out procedures.
Содержание SpectraSensors SS2100r
Страница 1: ...SS2100r H2S Analyzer Rack Mount Hardware Manual P N 4900002189 rev D...
Страница 2: ......
Страница 4: ......
Страница 8: ...SS2100r H2S Analyzer iv 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...SS2100r H2S Analyzer vi 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 18: ...SS2100r H2S Analyzer 1 8 4900002189 rev D 3 7 18 Figure 1 5 View of front and rear panels of the analyzer...
Страница 19: ...Introduction Hardware Manual 1 9 Figure 1 6 View of inside the analyzer enclosure with top panels removed...
Страница 25: ...Introduction Hardware Manual 1 15 Figure 1 10 Label placement on the inside of the analyzer enclosure...
Страница 34: ...SS2100r H2S Analyzer 2 8 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 44: ...SS2100r H2S Analyzer A 4 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...