NS 5.14 Firmware
2
–
34
4900002262 rev. B 3-23-21
Enter the required value.
Val Interval
The
Val Interval
parameter sets the number of days between autovalidation
cycles. The next scheduled validation cycle would occur in
Val Interval
days
at the
Val Start Time
. Enter the required value.
Val Perm Constant Kp
The
Val Perm Constant Kp
parameter is used for permeation validation
devices and defines the system constant that is determined at the factory
during calibration. Using the system constant, the permeation device can be
replaced with another permeation device with a different permeation rate, and
the correct new permeation concentration is calculated by the analyzer
software. The system constant K
p
will be constant over the life of the analyzer
provided the temperature, sample flow rate and pressure of the system are not
changed from the factory settings. If the system constant needs to be reset,
refer to
“Setting the Kp Value”
in the Hardware Manual for details on
recalculating the system constant.
Val Perm Rate Rp
The
Val Perm Rate Rp
parameter is used for permeation validation devices
and defines the permeation rate in ng/min referenced on the permeation
device certification. This certification is valid for a period of one (1) year.
However, the permeation device may be used longer than that period if a
factory certified validation concentration is not required. When the validation
<SET PARAMETER MODE>
Validation Wait Time
60
Enter a value (secs)
<SET PARAMETER MODE>
Val Interval
1
Enter a value (days)
<SET PARAMETER MODE>
Val Perm Constant Kp
0.24
0:Off >0:System Cons
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...