NS 5.14 Firmware
2
–
2
4900002262 rev. B 3-23-21
3.
The system goes through an initialization period while showing the
firmware version on the bottom line until the LCD displays the
Normal Mode screen.
4.
Enable
Peak Tracking
following the procedure outlined in
5.
After initialization and establishment of reference spectra, the LCD
displays four lines, the third of which is blank.
The measurements displayed are:
•
ANA
: Refers to the concentration in the sample cell (in units) of
the analyte/component selected in
Mode 2
.
•
P
: Pressure in the sample cell in units selected in
Mode 2
.
•
T
: Temperature in the sample cell in units selected in
Mode 2
.
6.
Continuous updates of the measurement parameters displaying on
the LCD indicates that the analyzer is operating normally.
Powering Down the Analyzer
It may be necessary to power down the analyzer for problem solving or
maintenance reasons. An approved switch or circuit breaker rated for 15 amps
should have been installed and clearly marked as the disconnecting device for
the analyzer.
For systems with heated enclosures, a
Temperature too Low
or
Temperature too High
fault will activate the
General Fault
Alarm
when the enclosure temperature is more than 5 °C above
or below the specified temperature. Once the enclosure has
reached the specified temperature, reset the
General Fault
Alarm
“To change parameters in Mode 2”
Definitions for the acronyms displayed on the LCD are given in the
section
<NORMAL MODE>
ANA: 4.0256ppmv
P: 954.4mb T: 76.1F
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...