
OXY5500 Optical Oxygen Analyzer
BA02195C
4
–
54
Hauser
The
Output
,
Mode
and
Error Trigger Level
settings will be applied to the
selected
Port
, which includes Port1, Port2 or the Input.
The
Output
of Port1 or Port2 can be Oxygen or Temperature.
The
Input
is always Pressure and cannot be changed.
The
Mode
of Port1 and Port2 can be set to one of the following:
•
Off: No input reading or output writing.
•
Linear: High and low value that has been set to correspond to 4mA
and 20ma. Values between these two settings will be calculated
linearly. Values outside this range will initiate the error trigger level.
•
Bilinear: High, mid and low value set to correspond to 4mA, 12mA
and 20mA respectively. This mode allows higher resolution in a
certain range. Refer to Figure 4–54 for an example.
Figure 4–53
Analogues - 4-20mA interface
settings
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...