BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
35
•
T0
: Temperature at the first calibration point.
– Select Auto to measure the temperature at the first calibration
point with the RTD probe (Pt100 temperature sensor).
– Select Manual if the first calibration point is known and remains
constant through the calibration process. Temperature values can
be input in
°
C,
°
F or K. Switch to the desired temperature unit
and change the temperature value in the input field.
•
T2nd
: Temperature at the second calibration point.
– Select
Auto at the first calibration point for automatic
temperature measurement.
– Select Manual to insert the changes to the calibration
temperature manually.
To proceed with the calibration, press the Next button at the top right of the
screen followed by the
.
Before starting the measurement, the OXY5500 must be calibrated. Refer to
the section
Calibrating the analyzer
Complete the calibration procedures in this section before starting the
measurement. First, refer to the required equipment and materials list in
Table 4–4. Figure 4–38 shows an illustration of the components to be used for
the cylinder regulator purge process.
Figure 4–37
Calibration temperature screen
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...