Replacing the Electro-Optical Module
Use the following procedure to replace and install the electro-optical module in
the OXY5500 analyzer.
Required tools and hardware
•
Flathead screwdriver
•
Philips screwdriver
•
Electro-optical module (P/N
EX0800000020)
Removing the electro-optical module
1.
Turn off power to the analyzer and open the enclosure door using a
standard flathead screwdriver to unlatch the lock.
2.
Disconnect the ribbon cable from the keypad and set aside.
3.
Disconnect the probes, power supply and pressure sensor from the
terminal blocks, as necessary. Refer to
1.
4.
Insert a flathead screwdriver into the clip extension at the top of the
electro-optical module as shown in Figure B–3.
Table B–1
Fuse specifications
Description
Rating
Cartridge fuse, 216 Series, 5 x 20 mm, Fast Act
800 mA, 250V
Photographs shown in this instruction are used to provide a
clearer illustration of the required steps only. DO NOT REMOVE
the base plate from the analyzer enclosure to complete this
instruction.
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...