OXY5500 Optical Oxygen Analyzer
BA02195C
5
–
18
Hauser
Ethernet subnet mask
Examples
Configuration of a continuous measurement
•
Precondition: Sensor is connected, and sensor constants and
calibration values are already set up correctly (OP-9).
The goal for this configuration is a continuous measurement with a 1-minute
interval with the pressure sensor and RTD (Pt100) are deactivated. Instead, a
manual fixed value will be transmitted. Refer to Table 5–17.
Table 5–16
Definition of register 5685
Start
Register
Number of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Write
Access
5685
8
Int32:
Ethernet
Subnet Mask
Octet 1
Int32:
Ethernet
Subnet
Mask
Octet 2
Int32:
Ethernet
Subnet
Mask
Octet 3
Int32:
Ethernet
Subnet
Mask
Octet 4
Yes
Table 5–17
Configuration for a continuous measurement
Step
Description
Register(s)
Value
1
Stop the measurement, if it is run
-
ning.
5707, 5708
0 (Int32)
2
Set pressure mode to “Manual”.
5705
0 (Boolean)
3
Set manual pressure to “1006.23”.
3147, 3148
1006.23 (Float)
4
Set temperature mode to “Manual”. 5611
0 (Boolean)
5
Set manual temperature to
“20.56”.
2409, 2410
20.56 (Float)
6
Set the interval rate to 1 min. (“60”
seconds).
3499, 3500
60 (Int32)
7
Activate the interval mode and
start the continuous measurement
immediately.
5707, 5708
5 (Int32 trans
-
lates to 00000101
binary)
Содержание SpectraSensors OXY5500
Страница 112: ...OXY5500 Optical Oxygen Analyzer BA02195C 5 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...OXY5500 Optical Oxygen Analyzer BA02195C A 4 Endress Hauser Figure A 2 Interconnect diagram AC ...
Страница 117: ...BA02195C OXY5500 Optical Oxygen Analyzer Endress Hauser A 5 Figure A 3 Interconnect diagram DC ...
Страница 120: ...OXY5500 Optical Oxygen Analyzer BA02195C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 142: ...OXY5500 Optical Oxygen Analyzer BA02195C B 22 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 145: ......
Страница 146: ...www endress com BA02195C 66 EN 01 21 ...