Prosonic S FMU95
Commissioning
E Hauser
63
"edit"
Use this parameter to enter, change or read a linearization table. There are the following options:
•
read:
The table editor is opened. The existing table can be read but not changed.
•
manual:
The table editor is opened. Table values can be entered and changed.
•
semi-automatic:
The table editor is opened. The level is automatically read by the Prosonic S. The measured value
(volume, weight or flow) must be entered by the user.
•
delete:
The linearization table is deleted.
The table editor
L00-FMU90xxx-19-00-00-de-006
"status table"
Use this parameter to enable or disable the linearization table.
Selection:
•
enabled
The table is used.
•
disabled
The table is
not
used. The measured values is transferred to the output without linearization.
No.
Level
Value
1
0,0000
0,0000
2
0,0000
0,0000
3
0,0000
0,0000
…
0,0000
0,0000
: go to next row
: go to previous row
: open marked row
for editing
No.
Level
Value
1
0,0000
0,0000
2
0,0000
0,0000
3
0,0000
0,0000
…
0,0000
0,0000
: navigate within the table
: (for "Level" und "Value”)
open marked number
for editing
: (for "No.")
open row editor
Row function
(before current row)
Query: new position
Return to the previous
step by this
key combination.
- Delete row
- Insert row
- Move row
Содержание Prosonic S FMU95
Страница 2: ......
Страница 104: ...Appendix Prosonic S FMU95 104 Endress Hauser...
Страница 107: ...Prosonic S FMU95 Index Endress Hauser 107...
Страница 108: ...www endress com worldwide BA00344F 00 EN 13 12 71164423 FM SGML 9 0 71164423...