Prosonic M FMU42

HART

Ultrasonic measuring technology

BA01963F/00/EN/01.22-00

71574096

2022-04-04

Содержание Prosonic M FMU42

Страница 1: ...Products Solutions Services Operating Instructions Prosonic M FMU42 HART Ultrasonic measuring technology BA01963F 00 EN 01 22 00 71574096 2022 04 04...

Страница 2: ...Prosonic M FMU42 HART 2 Endress Hauser Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555...

Страница 3: ...function of the operating menu 19 7 3 Access to the operating menu via the local display 20 7 4 Access to the operating menu via the operating tool 22 7 5 Locking or enabling parameter configuration...

Страница 4: ...lt in minor or medium injury NOTICE This symbol contains information on procedures and other facts which do not result in personal injury 1 2 2 Electrical symbols Ground connection Grounded clamp whic...

Страница 5: ...tructions document is located under the housing cover of the device The most important functions of the menu are summarized in this document Description of Device Parameters BA Reference manual for pa...

Страница 6: ...ations 2 4 Operational safety Risk of injury Operate the device only if it is in proper technical condition free from errors and faults The operator is responsible for ensuring failure free operation...

Страница 7: ...evice is installed and used as described in the Operating Instructions Provide additional protection for the device and data transfer to from the device IT security measures defined in the plant owner...

Страница 8: ...If one of these conditions is not satisfied contact your Sales Center 4 2 Product identification The device can be identified in the following ways Nameplate specifications Extended order code with br...

Страница 9: ...of the device to ensure that the device does not tilt or slip Follow the safety instructions and transport conditions for devices weighing more than 18 kg 39 6 lb IEC 61010 5 Mounting 5 1 Mounting re...

Страница 10: ...angle is defined as the beam angle At the energy density of the ultrasonic wave reaches half the value of the maximum energy density Ultrasonic waves are also emitted outside the signal beam and can...

Страница 11: ...he sensor membrane so that it is parallel to the surface of the water Observe the installation clearance of the flume or weir Enter the flow level Q h curve linearization curve via the FieldCare opera...

Страница 12: ...example BD Blocking distance of the sensor E Empty calibration enter during commissioning F Full calibration Hmax Maximum upstream level 5 1 5 Mounting examples A B A0038822 9 Mounting examples A Moun...

Страница 13: ...9 Blocking distance BD 0 4 m 1 3 ft Maximum range for liquids 10 m 33 ft Maximum range for bulk solids 5 m 16 ft 5 2 2 Blocking distance NOTICE If the blocking distance is undershot this may cause a...

Страница 14: ...ozzle diameter L Nozzle length Conditions Smooth interior surfaces in pipe nozzle Remove any edges or welded joints and burr on the inside of the tank side nozzle end To minimize disturbing factors no...

Страница 15: ...unting check Is the device free from damage visual inspection Does the measuring device comply with the measuring point specifications After aligning the housing is the process seal on the nozzle or f...

Страница 16: ...is fitted Remove the cable from the display 4 5 6 A0038243 A0038252 A0038253 Pull out the terminal module slightly using the pull loop Avoid moisture in the housing provide a loop to allow moisture to...

Страница 17: ...hield to the ground terminal in the connection compartment Tighten the cable gland 6 Close the housing 7 Switch on the power supply 6 3 Terminal assignment 2 4 3 1 2 3 4 L L I I 1 A0019269 13 Terminal...

Страница 18: ...age 14 to 30 V Current consumption 20 mA terminal voltage 8 to 30 V Ex d Current consumption 4 mA terminal voltage 14 to 30 V Current consumption 20 mA terminal voltage 11 to 30 V Fixed current termin...

Страница 19: ...upply voltage match the specifications on the nameplate No reverse polarity terminal assignment correct Do the cables used comply with the requirements Are the cable glands tightened correctly Is the...

Страница 20: ...display 7 3 1 Display 4 1 6 3 5 2 A0047443 15 Measured value display 1 Label e g measured value 2 Symbol 3 Value and unit e g 43 2 4 Position in menu e g 000 5 Bar graph 6 Position in menu The bar gra...

Страница 21: ...1 Operating keys Navigate up in the picklist Edit the numerical values within a function Navigate down in the picklist Edit the numerical values within a function Simultaneously Navigate to the left...

Страница 22: ...next character and continue editing until you have entered the entire value 3 If a symbol appears at the cursor press to accept the value entered Editing mode inactive 4 If a symbol appears at the cur...

Страница 23: ...Hardware locking only possible via the display press again simultaneously Hardware unlocking via communication is not possible 8 System integration 8 1 4 to 20 mA output with HART protocol 8 1 1 Opera...

Страница 24: ...View in the FieldCare operating program Select the function groups and functions of the device via the navigation bar Enter parameters in the input window Click the parameter names and call up exact...

Страница 25: ...r and labeling correct visual inspection Is the measuring device adequately protected against precipitation and direct sunlight 9 1 2 Post connection check Are the device or cables undamaged visual in...

Страница 26: ...with the local display 9 4 Configuring the measuring device 9 4 1 Basic setup F E BD SD A0038238 22 Parameters for the correct operation of the device BD Blocking distance SD Safety distance E Empty c...

Страница 27: ...e function A Dome ceiling B Horizontal cyl C Bypass or stilling well ultrasound guide pipe D No ceiling e g dumps open levels basins flume E Sphere F Flat ceiling Medium properties function 003 Select...

Страница 28: ...le measured value accurate measurement slow response time Turb surface For storage and buffer tanks with turbulent surfaces due to free filling mixing nozzles or small bottom agitators Emphasis on fil...

Страница 29: ...operation of the device D Distance sensor membrane product E Empty calibration zero point F Full calibration measuring span L Level BD Blocking distance SD Safety distance Empty calibration function 0...

Страница 30: ...el L Check whether the values displayed match the actual distance level Check distance function 051 Start the mapping process A B A0038449 26 Example of mapping A Distance too small B Distance Ok Sele...

Страница 31: ...vel The following situations can occur Distance and level correct Basic setup completed Distance and level incorrect perform another mapping in the Check distance 051 function Distance is correct but...

Страница 32: ...ally no interference echoes in front of the actual level signal Interference echoes must be below the mapping curve 1 2 3 4 5 6 7 A0038455 27 Envelope curve 1 Full calibration 2 Quality of the evaluat...

Страница 33: ...t 3 moves the curve to the left Activating vertical zoom mode 1 Press again 1 is displayed 2 increases the vertical scale 3 reduces the vertical scale The display symbol shows the current zoom factor...

Страница 34: ...e Output on alarm 010 function MAX 110 22 mA MIN 10 3 8 mA Hold the last value is held User specific value Warning W symbol flashes Device continues measuring error message is displayed Alarm warning...

Страница 35: ...61 Sensor temperature too high W103 Initializing If the message does not disappear replace the electronics W153 Initializing Wait a few seconds If the error remains active switch the voltage off on W6...

Страница 36: ...select the corresponding option for measurements in a bypass or ultrasound guide pipe 2 Perform mapping Error No change of measured value when filling or emptying Example 100 0 t A B A0038460 A Expect...

Страница 37: ...ct Turb surface or Add agitator 3 Increase the output damping 058 4 Select another installation position and or a larger sensor Error During filling or emptying the measured value drops Example 100 0...

Страница 38: ...nearization switches to linear table values are retained Switch the table in the Linearization 04 function group Linearization 041 function back on Navigation Diagnostics 0A function group Reset 0A3 f...

Страница 39: ...rom OS 4 6 with Rev 1 DD 1 V01 02 04 02 2004 Valid from documentation version BA237F 00 en 02 04 FMU42 added Software changes FMU42 added Compatible with HART Communicator DXR 375 V01 04 00 07 2006 Va...

Страница 40: ...nts Carry out repairs according to the instructions Only the Endress Hauser service team is permitted to modify a certified device and convert it to another certified version All repairs and modificat...

Страница 41: ...splay and operating unit FHX40 82 3 23 106 4 71 122 4 8 120 4 72 max 80 3 15 min 30 1 18 96 3 78 88 3 46 160 6 3 118 4 65 180 7 09 122 4 8 80 3 51 2 3 A B 4 150 5 91 8 5 0 3 3 6 3 0 25 1 A0038648 29 R...

Страница 42: ...information Brief Operating Instructions KA00202F 13 1 2 Weather protection cover 240 9 45 135 5 31 9 5 3 7 4 70 2 76 45 A0038977 30 Weather protection cover for F12 or T12 housing Unit of measurement...

Страница 43: ...ressures densities or other tank parameters The parameters are recorded and transmitted using the Fieldgate FXA42 gateway type This Web based software is installed on a local server and can also be vi...

Страница 44: ...4mA 0 D L F E BD SD A0038386 31 Parameters for the correct operation of the device BD Blocking distance SD Safety distance E Empty distance L Level D Distance from the sensor membrane to the surface...

Страница 45: ...14 to 30 V Current consumption 20 mA terminal voltage 8 to 30 V Ex d Current consumption 4 mA terminal voltage 14 to 30 V Current consumption 20 mA terminal voltage 11 to 30 V Fixed current terminal...

Страница 46: ...sured value resolution 2 mm 0 08 in 14 4 4 Measuring error Better than 4 mm 0 16 in or 0 2 of the measuring distance the larger value applies Observe typical measured error under reference operating c...

Страница 47: ...When housing is closed tested according to IP68 NEMA 6P 24 h at 1 83 m 6 ft under water IP66 NEMA 4X When housing is open tested according to IP20 NEMA 1 also degree of protection of the display 14 5...

Страница 48: ...larm 011 Outp echo loss 012 Ramp span min 013 Delay time 014 Safety distance 015 In safety dist 016 Acknowledge alarm 017 Temperature 03 Measured temp 030 Max temp limit 031 Max temperature 032 React...

Страница 49: ...0E1 Recording curve 0E2 Envelope curve display 0E3 Display 09 Language 092 Back to home 093 Format display 094 No of decimals 095 Sep character 096 Display test 097 Diagnostic 0A Present error 0A0 Pre...

Страница 50: ...nvelope curve Checking the measuring signal 31 Envelope curve display 32 Error codes 34 F Fieldcare Establishing the connection 26 FieldCare 23 Firmware history 39 Flow measurement 11 Full calibration...

Страница 51: ...Sensor features 13 Software locking 23 Start mapping 053 function 31 Storage transport 9 Structure of the operating menu 19 Supplementary documentation Additional documentation 5 Switching on Switchi...

Страница 52: ...www addresses endress com 71574096 71574096...