6 Commissioning
Prosonic Flow 92
42
EHauser
6.2.2
“Commissioning” Site Setup menu
This Site Setup menu guides you systematically through the setup procedure for all the
major device functions that have to be configured for standard measuring operation.
Use this “Site Setup” as well to determine the sensor distance or the wire length required
for sensor installation.
Fig. 33:
Site Setup menu for straightforward configuration of the major device functions
6.2.3
Sensor distance or wire length for installing the sensors
To get the sensor distance or wire length required for sensor installation, enter all of the
parameters displayed above for the “Site Data” function group. The result appears in
the bottom section of the screen display for the “Site Setup” group.
Use this value for:
Installing the Prosonic Flow W, P sensors, see Page 20 and 22
Installing the Prosonic Flow U sensors, see Page 23
!
Note!
You need the sensor distance for 2/4 traverse installation. You need the wire length for
single traverse installation.
F06-92x
x
x
x
x
x
-13-x
x
-x
x
-en-000