3.3
Installation instructions
3.3.1
Installing tensioning bands
For W sensor DN 50...4000
"
Caution!
Before use, check that the ratchet and springs are functioning correctly. Check tension-
ing bands for any damage.
Procedure for closing the tensioning band lock:
1.
Push one of the supplied threaded bolts on the tensioning band.
2.
Run the tensioning band around the pipe without twisting it and push the end
through the axis slit with the ratchet (a) open. Pretension manually by pulling the
free tensioning band end (a).
!
Note!
It is more difficult to release the tensioning bands if you do not pretension.
3.
Continuous tensioning via ratchet movement, i.e. move the lever back and forth (b)
until the tensioning band has the optimum tension.
4.
Then push down the lever (c).
"
Caution!
The tension lock (d) must engage on both sides!
Fig. 9:
Tensioning band lock
1 = Procedure for closing the tensioning band lock
2 = Procedure for opening the tensioning band lock
Procedure for opening the tensioning band lock:
1.
Pull back lever lock (e) and, at the same time, pull back the lever by 180° (f) until
the lever lock (g) is engaged. Remove the tensioning band.
F06-92xxxx
xx-17-xx
-06
-xx-000
Содержание prosonic flow 92
Страница 4: ...Brief operating instructions Prosonic Flow 92 4 Endress Hauser...
Страница 12: ...2 Identification Prosonic Flow 92 12 Endress Hauser...
Страница 26: ...3 Installation Prosonic Flow 92 26 Endress Hauser...
Страница 34: ...4 Wiring Prosonic Flow 92 34 Endress Hauser...
Страница 40: ...5 Operation Prosonic Flow 92 40 Endress Hauser...
Страница 46: ...7 Maintenance Prosonic Flow 92 46 Endress Hauser...
Страница 96: ...Index Prosonic Flow 92 96 Endress Hauser...