9 Trouble-shooting
Hauser
85
9.4
Process errors without messages
Symptoms
Remedial measures
Note:
You may have to change or correct certain settings in functions in the matrix in order to rectify faults. The functions out-
lined below, such as DISPLAY DAMPING, for example, are described in detail in the “Description of Device Functions”
manual.
Flow values are negative, even
though the fluid is flowing
forwards through the pipe.
1.
Check wiring
If necessary, reverse the connections at terminals “up” and “down”.
2.
Change the setting in the “INSTALLATION DIRECTION, SENSOR” function
accordingly
Measured value reading fluctu-
ates even though flow is steady.
1.
Check the fluid for presence of gas bubbles.
2.
“TIME CONSTANT” function (current output)
→
Increase value
3.
“DISPLAY DAMPING” function
→
Increase value
There are differences between
the flowmeter's internal totalizer
and the external metering
device.
This symptom is due primarily to backflow in the piping, because the pulse output
cannot subtract in the “STANDARD or SYMMETRY” measuring modes.
The following solution is possible:
Allow for flow in both directions.
Measured value reading shown
on display, even though the fluid
is at a standstill and the measur-
ing tube is full.
1.
Check the fluid for presence of gas bubbles.
2.
Activate the “LOW FLOW” function, i.e. enter or increase the value for the
switching point.
The current output signal is
always 4 mA, irrespective of the
flow signal at any given time.
1.
Select the “BUS ADDRESS” function and change the setting to “0”.
2.
Low flow cut off too high. Reduce the corresponding value in the “LOW FLOW
CUTOFF” functions (ON-/OFF-VALUE).
The fault cannot be rectified or
some other fault not described
above has occurred.
In these instances, please contact
your Hauser service
organisation.
The following options are available for tackling problems of this nature:
Request the services of an Hauser service technician
If you contact our service organisation to have a service technician sent out,
please be ready with the following information:
– Brief description of the fault
– Nameplate specifications (Page 9 ff.): Ordering code and serial number
Returning devices to Hauser
The procedures on Page 8 must be carried out before you return a flowmeter
requiring repair or calibration to Hauser.
In all cases, enclose a fully completed “Declaration of contamination” form with the
flowmeter. You will find a preprinted form at the end of these Operating Instruc-
tions.
Replace transmitter electronics
Components in the measuring electronics defective
→
Order spare
part
→
Содержание Prosonic Flow 90
Страница 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Страница 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Страница 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Страница 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Страница 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Страница 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Страница 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Страница 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Страница 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...