Commissioning
Proservo NMS83
78
Hauser
9.3.4
Reference calibration
The reference calibration defines the zero distance position of the displacer from the
mechanical stop.
1. Navigate to: Setup → Calibration → Reference calibration → Reference calibration
2. Select the
Start
option
3. Check the reference position (e.g. 70 mm (2.76 in)).
The reference position is preset prior to delivery.
4. Confirm that the displacer is correctly attached to the measuring wire.
5. The reference calibration starts automatically.
This completes the reference calibration.
1
1.
2.
3.
R
A0028003
37
Reference calibration sequence
1
Mechanical stop
R
Reference position
Содержание Proservo NMS83
Страница 173: ...Proservo NMS83 Operating menu Endress Hauser 173 Additional information Read access Operator Write access Maintenance ...
Страница 182: ...Operating menu Proservo NMS83 182 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 218: ...Operating menu Proservo NMS83 218 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 239: ...Proservo NMS83 Operating menu Endress Hauser 239 Additional information Read access Operator Write access ...
Страница 260: ...Operating menu Proservo NMS83 260 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 275: ...Proservo NMS83 Operating menu Endress Hauser 275 Additional information Read access Operator Write access Maintenance ...
Страница 278: ...Operating menu Proservo NMS83 278 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 310: ...Operating menu Proservo NMS83 310 Endress Hauser 15 4 4 Device check submenu Navigation Diagnostics Device check ...
Страница 317: ......
Страница 318: ...www addresses endress com 71318331 71318331 ...