Proline t-mass A 150 HART
Electrical connection
Hauser
23
7
Electrical connection
7.1
Connection conditions
7.1.1
Required tools
• For cable entries: Use corresponding tools
• Wire stripper
• When using stranded cables: Crimping tool for wire end ferrule
• Flat blade screwdriver≤ 3 mm (0.12 in)
7.1.2
Requirements for connecting cable
The connecting cables provided by the customer must fulfill the following requirements.
Electrical safety
In accordance with applicable federal/national regulations.
Cable specification
Permitted temperature range:
• –40 °C (–40 °F)...≥ 80 °C (176 °F)
• Minimum requirement for cable temperature range: ambient tempe20 K
Current output
For 4-20 mA HART: Shielded cable recommended. Observe grounding concept of the
plant.
Pulse/frequency/switching output
Standard installation cable is sufficient
Cable diameter
• Included cable glands: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• Core cross-sectional area0.5 to 1.5 mm
2
(21 to 16 AWG)
7.1.3
Requirements for the supply unit
Device supply voltage
DC 24 V (18 to 30 V)
The power supply circuit must comply with ELV (BS 7671) requirements.
Pulse/frequency/status supply voltage
An external power supply is required for each output.
Order characteristic for
"Output"
Maximum terminal voltage
Option
B, K
DC 30 V
Load
0 to 750 Ω, depending on the external supply voltage of the power supply unit