Proline t-mass 65 PROFIBUS DP/PA
Technical data
Hauser
117
Media
The following media and media mixtures can be measured. A mixture can consist of up to 8
components from the following list.
Note
Other media (e.g. ozone) available on request. Contact your Hauser sales center for
clarification.
Pressure-temperature
ratings
!
Note!
An overview of the pressure-temperature ratings for the process connections is provided in
the Technical Information
Flow limit
111.
The velocity in the measuring tube should not exceed 130 m/s (427 ft/s), (in air).
Pressure loss
Negligible (without flow conditioner).
Refer to Applicator for the precise calculation
Medium pressure range
(nominal pressure)
t-mass F:
–0.5 to 40 barg (–7.25 to 580 psig)
t-mass I:
–0.5 to 20 barg (–7.25 to 290 psig)
Limiting medium pressure
range (nominal pressure)
The material load diagrams (pressure-temperature ratings) for the process connections are
provided in the separate "Technical Information" document for the particular measuring
device. This can be downloaded as a PDF file at www.endress.com.
A list of the "Technical Information" documents:
Hot tap, process pressure
The Hot tap for mounting and removal at process pressure may only be used with non-toxic,
innocuous gases classified as "Group II" in accordance with European Directive 67/548/EEC
Art. 2.
Medium pressure version
Max. process pressure: 20 barg (290 psig)
Max. extraction press: 16 barg (230 psig)
Max. extraction temperature: +50 °C (+122 °F)
Min. sensor length: 435 mm (17 in)
Low pressure version
Max. process pressure: 20 barg (290 psig)
Max. extraction press: 4,5 barg (65 psig)
Max. extraction temperature: +50 °C (+122 °F)
Min. sensor length: 335 mm (13 in)
AIR
AMMONIA
ARGON
BUTANE
CARBON DIOXIDE
CARBON MONOXIDE
CHLORINE
ETHANE
ETHYLENE
HELIUM 4
HYDROGEN NORMAL
HYDROGEN CHLORIDE
HYDROGEN SULFIDE
KRYPTON
METHANE
NEON
NITROGEN
OXYGEN
PROPANE
XENON