Proline Prosonic Flow P
Installation
Hauser
19
7. Push down the tensioning screw and tighten the strapping bands so they cannot slip.
1
A0043372
7
Tightening the tensioning screws of the strapping bands
1
Sensor holder
8. If necessary, shorten the strapping bands and trim the cut edges.
L
WARNING
Risk of injury!
‣
To avoid sharp edges, trim the cut edges after shortening the strapping bands. Wear
suitable gloves and protective goggles.
The visible measuring pipe surface must be clean (free from flaking paint and/or rust) to
ensure good acoustic contact.
Sensor holder with strapping bands (medium nominal diameters)
Can be used for
• Measuring devices with measuring range DN 50 to 4000 (2 to 160")
• Mounting on pipes DN ≤ 600 (24")
Procedure:
1. Fit the mounting bolt over strapping band 1.
2. Position strapping band 1 as perpendicular as possible to the measuring pipe axis
without twisting it.
3. Guide the end of strapping band 1 through the strapping band lock.
4. Tighten strapping band 1 as tightly as possible by hand.