Proline Prosonic Flow 93C FOUNDATION Fieldbus
Technical data
Hauser
81
Cable specifications
Only use the connecting cables supplied by Hauser.
Different versions of the connecting cables are available
Prosonic Flow W
• Cable material made of PVC (standard) or PTFE (for higher temperatures)
• Cable length: 5 to 60 m (16.4 to 196.8 ft)
!
Note!
To ensure correct measuring results, route the connecting cable well clear of electrical machines
and switching elements.
Power consumption
AC: < 18 VA (incl. sensor)
DC: < 10 W (incl. sensor)
Switch-on current
:
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
Power supply failure
Lasting min. 1 power cycle:
EEPROM and T-DAT save measuring system data if the power supply fails
Potential equalization
For potential equalization, no special measures are necessary. With regard to devices suitable for
potentially explosive atmospheres, please refer to the information in the Ex-specific supplementary
documentation.
10.1.6
Performance characteristics
Reference operating
conditions
• Fluid temperature: +20 to +30 °C
• Ambient temperature: +22 °C ± 2 K
• Warm-up period: 30 minutes
• Sensors and transmitter are grounded.
• The measuring sensors are correctly installed.
Maximum measured error
At a flow velocity of > 0.3m/s (1 ft/s) and a Reynolds number > 10000, the following error limits
of the measuring device are guaranteed:
!
Note!
The Prosonic Flow 93C Inline sensor is also available without factory flow calibration. The error
limits without calibration are ±1.5 % o.r. ± 3 mm/s
Repeatability
±0.3 % for flow velocity > 0.3 m/s (1 ft/s)
Nominal diameter
Guaranteed error limits of the device
Report
DN 300 to 2000 (12 to 80")
±0.5 % o.r. ± 3 mm/s
Factory measurement report
o.r. = of reading