Operation options
Proline Promass O 100 HART
38
Hauser
Enabling the Web server
If the Web server is disabled it can only be re-enabled with the
Web server functionality
parameter via the following operating options:
Via "FieldCare" operating tool
8.3.7
Logging out
Before logging out, perform a data backup via the
Data management
function
(upload configuration from device) if necessary.
1. Select the
Logout
entry in the function row.
The home page with the Login box appears.
2. Close the Web browser.
3. Reset the modified properties of the Internet protocol (TCP/IP) if they are no longer
8.4
Access to the operating menu via the operating tool
8.4.1
Connecting the operating tool
Via HART protocol
This communication interface is present in the following device version:
Order code for "Output", option
B
: 4-20 mA HART, pulse/frequency/switch output
1
2
3
5
7
4
6
A0016948
11
Options for remote operation via HART protocol
1
Control system (e.g. PLC)
2
Field Communicator 475
3
Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM)
4
Commubox FXA195 (USB)
5
Field Xpert SFX350 or SFX370
6
VIATOR Bluetooth modem with connecting cable
7
Transmitter