Measuring range
Measuring ranges for liquids
DN
Measuring range full scale values
min(F)
to
max(F)
[mm]
[in]
[kg/h]
[lb/min]
8
³⁄₈
0 to 2
000
0 to 73.50
15
½
0 to 6
500
0 to 238.9
15 FB
½ FB
0 to 18
000
0 to 661.5
25
1
0 to 18
000
0 to 661.5
25 FB
1 FB
0 to 45
000
0 to 1
654
40
1½
0 to 45
000
0 to 1
654
40 FB
1½ FB
0 to 70
000
0 to 2
573
50
2
0 to 70
000
0 to 2
573
50 FB
2 FB
0 to 180
000
0 to 6
615
80
3
0 to 180
000
0 to 6
615
FB = Full bore
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the formula below:
max(G)
=
max(F)
· ρ
G
: x
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]
max(G)
<
max(F)
max(G)
can never be greater than
max(F)
ρ
G
Gas density in [kg/m³] at operating conditions
DN
x
[mm]
[in]
[kg/m
3
]
8
³⁄₈
60
15
½
80
15 FB
½ FB
90
25
1
90
25 FB
1 FB
90
40
1½
90
40 FB
1½ FB
90
50
2
90
50 FB
2 FB
110
80
3
155
110
FB = Full bore
To calculate the measuring range, use the
Applicator
sizing tool (→ 82)
Calculation example for gas
• Sensor: Promass I, DN 50
• Gas: Air with a density of 60.3 kg/m³ (at 20 °C and 50 bar)
• Measuring range (liquid):70
000 kg/h
• x = 90 kg/m³ (for Promass I, DN 50)