Proline Promass E 300 Modbus RS485
System integration
Hauser
69
9
System integration
9.1
Overview of device description files
9.1.1
Current version data for the device
Firmware version
01.05.zz
• On the title page of the Operating Instructions
• On the transmitter nameplate
• Firmware version
Diagnostics → Device information → Firmware
version
Release date of firmware version
08.2019
---
For an overview of the different firmware versions for the device → 160
9.1.2
Operating tools
The suitable device description file for the individual operating tools is listed in the table
below, along with information on where the file can be acquired.
Operating tool via service
interface (CDI) or Modbus
interface
Sources for obtaining device descriptions
FieldCare
•
→ Download Area
• CD–ROM (contact Hauser)
• DVD (contact Hauser)
DeviceCare
•
→ Download Area
• CD–ROM (contact Hauser)
• DVD (contact Hauser)
9.2
Compatibility with earlier model
If the device is replaced, the measuring device Promass 300 supports the compatibility of
the Modbus registers for the process variables and the diagnostic information with the
previous model Promass 83. It is not necessary to change the engineering parameters in
the automation system.
Compatible Modbus registers: process variables
Process variable
Compatible Modbus registers
Mass flow
2007
Volume flow
2009
Corrected volume flow
2011
Density
2013
Reference density
2015
Temperature
2017
Totalizer 1
2610
Totalizer 2
2810
Totalizer 3
3010