10.3.4 Configuring the low flow cut off
The
Low flow cut off
submenu contains parameters that have to be set for configuring the
low flow cut off.
Navigation path
"Setup" menu → Low flow cut off
Structure of the submenu
Low flow cut off
→
Assign process variable
On value low flow cut off
Off value low flow cutoff
Pressure shock suppression
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection/
User entry
Factory
setting
Assign process
variable
–
Select the process
variable for low flow cut
off.
• Off
• Mass flow
• Volume flow
• Corrected
volume flow
Mass flow
On value low
flow cut off
One of the following options is
selected in the
Assign process
variable
parameter:
• Mass flow
• Volume flow
• Corrected volume flow
Enter the on value for
low flow cut off.
Max. 15-digit,
positive floating-
point number
For liquids:
depends on
country and
nominal
diameter
Off value low
flow cut off
One of the following options is
selected in the
Assign process
variable
parameter:
• Mass flow
• Volume flow
• Corrected volume flow
Enter the off value for
low flow cut off.
0 to 100 %
50 %
Pressure shock
suppression
One of the following options is
selected in the
Assign process
variable
parameter:
• Mass flow
• Volume flow
• Corrected volume flow
Enter the time interval
for signal suppression (=
active pressure shock
suppression).
0 to 100 s
0 s