Commissioning Instructions for PTB gas approval Proline Promass 84
Functions relevant to custody transfer
Hauser
37
7.24
Basic function
Process parameter
Configuration
7.25
Basic function
Process para.
EPD parameter
7.26
Basic function
Process para.
Ref. parameter
Function name
No.
Description
Configuration
ASSIGN LOW FLOW CUT
OFF
6400 Use this function to assign a switching function to the
low flow cut off.
Options:
e.g. MASS FLOW
ON-VALUE LOW FLOW CUT
OFF
6402 Use this function to enter the value for the switch-on
point.
User input:
e.g. 0 kg
OFF-VALUE LOW FLOW
CUTOFF
6403 Use this function to enter the value for the switch-off
point.
User input:
e.g. 2 kg
PRESSURE SHOCK
SUPPRESSION
6404 Use this function to specify the duration of pressure
shock suppression.
User input:
e.g. 0.0 s
Function name
No.
Description
Configuration
EMPTY PIPE DETECTION
6420 Use this function to activate empty pipe detection.
Options:
e.g. OFF
EPD VALUE LOW
6423 Use this function to enter the lower density value for
fluid monitoring.
User input:
e.g. 0.2000 kg/l
EPD VALUE HIGH
6424 Use this function to enter the upper density value for
fluid monitoring.
User input:
e.g. 6.0000 kg/l
EPD RESPONSE TIME
6425 Use this function to specify the time period which must
elapse before a notice or fault message is generated.
User input:
e.g. 1.0 s
EPD EXC.CURR.
6426 Use this function to specify the excitation current which
generates a notice or fault message.
User input:
e.g. 100 mA
Function name
No.
Description
Configuration
CORRECTED VOLUME
CALCULATION
6460 Use this function to select how the standard volume
flow is to be calculated.
Options: e.g. FIXED
REFERENCE DENSITY
FIXED REFERENCE DENSITY 6461 Use this function to enter the density value for
calculating the standard volume flow.
User input:
e.g. 1 kg/Nl
EXPANSION COEFFICIENT
6462 Use this function to enter the expansion coefficient used
to calculate the temperature-compensated density
functions.
User input:
e.g. 0.5000 e
–3
[1/K]
EXPANSION COEFFICIENT
SQUARE
6463 Use this function to enter the squared expansion
coefficient in case temperature compensation is not
linear.
User input:
e.g. 0 e
–6
[1/K
2
]
REFERENCE TEMPERATURE 6464 Use this function to enter the reference temperature for
calculating: corrected vol. flow, corrected vol. and
reference density.
User input:
e.g. 20.000 °C