3 Installation
Proline Promass 83 FOUNDATION Fieldbus
16
Hauser
3.2.3
Orientation
Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of
flow (direction in which the fluid flows through the pipe).
Orientation Promass A
Vertical:
Recommended orientation with upward direction of flow. When fluid is not flowing, entrained
solids will sink down and gases will rise away from the measuring tube.
The measuring tubes can be completely drained and protected against solids build-up.
Horizontal:
When installation is correct the transmitter housing is above or below the pipe. This arrangement
means that no gas or solid deposits can accumulate in the curved measuring tube (single-tube
system).
Do not install the sensor in such a way that it is suspended in the pipe, in other words without sup-
port or attachment. This is to avoid excessive strain at the process connection. The base plate of the
sensor housing is designed for mounting on a tabletop, wall or post.
F06-xxxxxxxx-11-00-00-xx-005
Fig. 7:
Vertical and horizontal orientation (Promass A)
Содержание Proline Promass 83
Страница 24: ...3 Installation Proline Promass 83 FOUNDATION Fieldbus 24 Endress Hauser ...
Страница 34: ...4 Wiring Proline Promass 83 FOUNDATION Fieldbus 34 Endress Hauser ...
Страница 64: ...6 Commissioning Proline Promass 83 FOUNDATION Fieldbus 64 Endress Hauser ...
Страница 66: ...7 Maintenance Proline Promass 83 FOUNDATION Fieldbus 66 Endress Hauser ...
Страница 90: ...9 Trouble shooting Proline Promass 83 FOUNDATION Fieldbus 90 Endress Hauser ...
Страница 116: ...Index Proline Promass 83 FOUNDATION Fieldbus 116 Endress Hauser ...
Страница 118: ...www endress com worldwide BA065D 06 en 12 04 50098623 FM SGML 6 0 ...