Proline Promag P 100
Electrical connection
Hauser
29
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)
7.3
Hardware settings
7.3.1
Setting the device address
EtherNet/IP
The IP address of the measuring device can be configured for the network via DIP switches.
Addressing data
IP address and configuration options
1st octet
2nd octet
3rd octet
4th octet
192.
168.
1.
XXX
↓
↓
Can only be configured via software addressing
Can be configured via
software addressing
and hardware
addressing
IP address range
1 to 254 (4th octet)
IP address broadcast
255
Addressing mode ex works
Software addressing; all DIP switches for hardware addressing are set to OFF.
IP address ex works
DHCP server active
For device addressing via software (→ 42)
Setting the address
OFF ON
1
2
3
4
5
6
7
8
9
10
1
2
4
8
16
32
64
128
- Write protection
- Default Ethernet
network settings
IP 192.168.1.212
IP
Address setting
(last octet)
A0017913