Electrical connection
Proline Promag L 800
42
Hauser
7.5 Potential equalization
#
Warning!
Integrate the measuring system into the potential equalization system .
7.5.1 Potential equalization requirements
Please consider the following to ensure correct measurement:
• The fluid and sensor have the same electrical potential
• Company-internal grounding concepts
• Material and grounding of the pipes
7.5.2 Connection examples for potential equalization
Connection example in standard situations
Metal, grounded pipe
A0016315
Abb. 25: Potential equalization via Measuring tube
Connection example in special situations
Plastic pipe or pipe with insulating lining
This connection method is also to be used when:
• Potential equalization is not customary
• Equalizing currents are present
A0016318
Abb. 26: Potential equalization via ground terminal and ground disk