Commissioning
ProlineйPromagйDй400йHART
96
Hauser
Parameters which can always be modified via the local display
Certainйparametersйthatйdoйnotйaffectйtheйmeasurementйareйexceptedйfromйwriteйprotection
viaйtheйlocalйdisplay.йDespiteйtheйdefinedйaccessйcode,йtheseйparametersйcanйalwaysйbe
modifiedйevenйifйtheйotherйparametersйareйlocked.
Parametersйforйconfiguring
theйlocalйdisplay
Parametersйforйconfiguring
theйtotalizer
Language
Formatйdisplay
ControlйTotalizer
Contrastйdisplay
Presetйvalue
Displayйinterval
Resetйallйtotalizers
Defining the access code via the Web browser
1. Navigateйtoйtheй
Enter access code
йparameter.
2. Defineйaйmax.й4-digitйnumericйcodeйasйanйaccessйcode.
3. Enterйtheйaccessйcodeйagainйtoйconfirmйtheйcode.
TheйWebйbrowserйswitchesйtoйtheйloginйpage.
Ifйnoйactionйisйperformedйforй10йminutes,йtheйWebйbrowserйautomaticallyйreturnsйto
theйloginйpage.
TheйuserйroleйwithйwhichйtheйuserйisйcurrentlyйloggedйonйviaйtheйWebйbrowserйis
indicatedйbyйtheй
Access status tooling
йparameter.йNavigationйpath:йOperationй
Accessйstatusйtooling
10.7.2 Write protection via write protection switch
Unlikeйwriteйprotectionйviaйuser-specificйaccessйcode,йthisйallowsйwriteйaccessйtoйtheйentire
operatingйmenuй-йotherйthanйtheй
"Contrast display" parameter
й-йtoйbeйlocked.
Theйparameterйvaluesйareйnowйreadйonlyйandйcannotйbeйeditedйanyйmoreйсexception
"Contrast display" parameter
):
• Viaйlocalйdisplay
• ViaйserviceйinterfaceйсCDI-RJ45)
• ViaйHARTйprotocol
ON
+ - + -
+ - + -
OFF
ййA0017260
1. Loosenйtheй4йfixingйscrewsйonйtheйhousingйcoverйandйopenйtheйhousingйcover.
2. SettingйtheйwriteйprotectionйswitchйсWP)йonйtheйmainйelectronicsйmoduleйtoйtheйON
positionйenablesйtheйhardwareйwriteйprotection.йSettingйtheйwriteйprotectionйswitch
сWP)йonйtheйmainйelectronicsйmoduleйtoйtheйOFFйpositionйсfactoryйsetting)йdisables
theйhardwareйwriteйprotection.