3 Transducer Block
Device functions Proline Promag 55 FOUNDATION Fieldbus
96
Hauser
Config. -
Display Damping
AUTO - OOS
Use this parameter to enter a time constant defining how the display reacts
to severely fluctuating flow variables, either very quickly (enter a low time
constant) or with damping (enter a high time constant).
User input:
0 to 100 seconds
Factory setting:
1 s
!
Note!
Setting the time constant to zero seconds switches off damping.
Config. -
Contrast LCD
AUTO - OOS
Use this parameter to optimize display contrast to suit local operating
conditions.
User input:
10 to 100%
Factory setting:
50%
Config. -
Backlight
AUTO - OOS
Use this parameter to optimize the backlight to suit local operating
conditions.
User input:
0 to 100%
!
Note!
Entering the value "0" means that the backlight is "switched off". The
display then no longer emits any light, i.e. the display texts can no longer
be read in the dark.
Factory setting:
50%
Operation -
Test Display
AUTO - OOS
Use this parameter to test the operability of the local display and its pixels.
Options:
ON
OFF
Factory setting:
OFF
Test sequence:
1.
Start the test by selecting ON.
2.
All pixels of the main line, additional line and information line are
darkened for minimum 0.75 seconds.
3.
Main line, additional line and information line show an "8" in each
field for minimum 0.75 seconds.
4.
Main line, additional line and information line show a "0" in each field
for minimum 0.75 seconds.
5.
Main line, additional line and information line show nothing (blank
display) for minimum 0.75 seconds.
When the test is completed, the local display returns to its initial state.
"Display" Transducer Block/base index 1800:
Parameter
Write access with
operating mode
(MODE_BLK)
Description
Содержание Proline Promag 55
Страница 60: ...8 Block SUPERVISION Device Functions Proline Promag 55 FOUNDATION Fieldbus 60 Endress Hauser ...
Страница 66: ...10 Index local operation Device Functions Proline Promag 55 FOUNDATION Fieldbus 66 Endress Hauser ...
Страница 68: ...Table of contens FOUNDATION Fieldbus Device Functions Proline Promag 55 FOUNDATION Fieldbus 68 Endress Hauser ...
Страница 129: ...Device Functions Proline Promag 55 FOUNDATION Fieldbus Endress Hauser 129 ...
Страница 130: ...www endress com worldwide BA127D 06 en 09 06 71032705 FM SGML 6 0 ...