Proline Promag 55 PROFIBUS PA
Wiring
Hauser
55
4.6
Degree of protection
The devices fulfill all the requirements for IP 67 (NEMA 4X).
Compliance with the following points is mandatory following installation in the field or servicing,
in order to ensure that IP 67 protection (NEMA 4X) is maintained:
• The housing seals must be clean and undamaged when inserted into their grooves. The seals must
be dried, cleaned or replaced if necessary.
• All screws and screw covers must be firmly tightened.
• The cables used for connection must be of the specified external diameter
• Tighten cable glands to prevent leakages.
• The cables must loop down before they enter the cable entries ("water trap"). This arrangement
prevents moisture penetrating the entry. Always install the measuring device in such a way that
the cable entries do not point up.
• Close off unused cable entries using suitable insert plugs.
• Do not remove the grommet from the cable entry.
a0001914
Fig. 40:
Installation instructions, cable entries
"
Caution!
Do not loosen the threaded fasteners of the sensor housing, as otherwise the degree of protection
guaranteed by Hauser no longer applies.
!
Note!
The Promag S sensor can be supplied with IP 68 rating (permanent immersion in water to a depth
of 3 meters). In this case the transmitter must be installed remote from the sensor.
a