6 Commissioning
Proline Promag 51
78
Hauser
#
Warning!
• Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.
• Wait at least 10 minutes for heat to dissipate before opening an Ex-rated device.
1.
Switch off power suppy.
2.
Remove the I/O board
→
3.
Position the jumper according to Fig. 46.
"
Caution!
Risk of destroying the measuring device. Set the jumpers exactly as shown in Fig. 46. Incor-
rectly set jumpers can cause overcurrents that would destroy either the measuring device or
external devices connected to it.
4.
Installation of the I/O board is the reverse of the removal procedure.
A0001044
Fig. 44:
Configuring the current output (I/O board)
1
Active current output (Factory setting)
2
Passive current output
6.5
Data storage device (HistoROM)
At Hauser, the term HistoROM refers to various types of data storage modules on which
process and measuring device data are stored.
6.5.1
HistoROM/S-DAT (sensor-DAT)
The S-DAT is an exchangeable data storage device in which all sensor relevant parameters are
stored, i.e., diameter, serial number, calibration factor, zero point.
1
2
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...